Oxygen Regulator Instructions

- Remove plastic cover from cylinder neck.
- Make sure there is a washer on the largest pin of the regulator or the oxygen conserving device (OCD).
- Attach regulator or an oxygen conserving device (OCD) by sliding it onto the cylinder neck until 3 pins line up with 3 holes on cylinder (see figure 1).
- Hand, tighten screw on regulator or the oxygen conserving device (OCD). (see figure 2)
- Open cylinder by turning valve counterclockwise until pressure gauge on regulator or the oxygen conserving device (OCD) indicates how full tank is. (see figure 3)
- Connect tubing and/or nasal cannula securely to nipple on regulator or the oxygen conserving device (OCD).
- Set the flow setting to your prescribed setting by turning the knob on the regulator or the oxygen conserving device (OCD). (see figure 4)
- While in use, periodically check the contents gauge. To make sure you have enough to get to your destination and back.
- Close cylinder valve before removing regulator or the oxygen conserving device (OCD). Bleed off the remaining oxygen before removing regulator or the oxygen conserving device (OCD). This can sometimes create a load hissing noise.
- Turn on flow knob again after cylinder valve is closed to release remaining oxygen. Turn flow knob off before removing regulator or the oxygen conserving device (OCD).